Subaortic pannus causing complete outlet obstruction after bioprosthetic aortic valve replacement in a patient with left ventricular assist device: a case report.


ABSTRACT:

Background

Subaortic pannus formation complicates bioprosthetic aortic valve (AV) replacement. We report an extreme case in a continuous-flow left ventricular assist device (LVAD) patient.

Case summary

A 49-year-old Caucasian female with dilated cardiomyopathy was bridged to transplant with a HeartWare Ventricular Assist Device (Medtronic). Duration of support was prolonged, 6 years & 7 months, due to allosensitization requiring desensitization. Pump thrombosis occurred 2 years & 4 months post-LVAD requiring alteplase thrombolysis. The patient underwent bioprosthetic AV replacement 3 years & 10 months post-LVAD for symptomatic aortic incompetence. Transthoracic echocardiography (TTE) performed 1 year and 2 years post-bioprosthetic AV replacement repeatedly demonstrated an AV