Evaluation of cryptococcal antigen testing using a novel chemiluminescence assay in two medical centers of China.


ABSTRACT:

Objective

This study aimed to assess the efficacy of innovative Chemiluminescence Immunoassay (CLIA) in testing Cryptococcal Antigen (CrAg) across two medical centers, employing the FDA-approved CrAg Lateral Flow Assay (LFA) by IMMY as a reference standard.

Methods

The study encompassed patients diagnosed with cryptococcosis at West China Hospital of Sichuan University (HX) between July 2022 and May 2023, and Suzhou Fifth People's Hospital (SZ) from September 2020 to September 2023. All specimens underwent simultaneous detection using the LFA (IMMY, Norman, USA) and CLIA (Chuanglan, Suzhou, China).

Results

A total of 628 patients were enrolled, revealing a remarkable 99.20% concordance between LFA and CLIA (623/628, 99.20%). The LFA exhibited a sensitivity of 96.83% (