Quantification of safe operation conditions for large-area platinum-iridium electrodes in neurostimulation application.


ABSTRACT:

Objective

Using electrochemical characterization methods of stimulation electrodes as well as accelerated stimulation examinations, a safe operating field for stimulation is investigated for particularly very large Pt-Ir macroelectrodes in a Laplace configuration.

Approach

Traditional methods such as Electrochemical Impedance Spectroscopy, Cyclic Voltammetry and biphasic, charge balanced current pulses were applied on Pt-Ir macroelectrodes in phosphate buffered saline solution to investigate reversible boundaries. These experiments were adapted to approach realistic working conditions.

Main results

Investigating operational conditions close to realistic use cases have shown an anti-correlation between higher scan rates and the occurrence of irreversible reactions. In