ABSTRACT: Background
SARS-CoV-2 antigen rapid detection tests (RDTs) emerged as point-of-care diagnostics alongside reverse transcription polymerase chain reaction (RT-qPCR) as reference.Methods
In a prospective performance assessment from 12 November 2020 to 30 June 2023 at a single centre tertiary care hospital, the sensitivity and specificity (primary endpoints) of RDTs from three manufacturers (NADAL®, Panbio™, MEDsan®) were compared to RT-qPCR as reference standard among patients, accompanying persons and staff aged ≥ six month in large-scale, clinical screening use. Regression models were used to assess influencing factors on RDT performance (secondary endpoints).Findings
Among 78,798 paired RDT/RT-qPCR results analysed, overall RDT sensitivity was 34.5% (695/2016; 95%