ABSTRACT: Background and purpose
Conventionally, the quality of radiotherapy treatment plans is assessed through visual inspection of dose distributions and dose-volume histograms. This study developed a framework to evaluate plan quality using dose, complexity, and robustness metrics. Additionally, a method for predicting plan robustness metrics using dose and complexity metrics was introduced for cases where plan robustness evaluation is unavailable or impractical.Materials and methods
The framework and prediction models were developed and validated using 103-bronchial Volumetric Modulated Arc Therapy (VMAT)-plans. The application of the framework was demonstrated using 25-VMAT-plans. To identify significant metrics for plan evaluation, 122-metrics were analysed and narrowed down u