Multicenter Evaluation of the QIAstat-Dx and the BioFire Multiplex Panel Tests for the Detection of Respiratory Pathogens.
Ontology highlight
ABSTRACT: Syndromic multiplex panel testing enables simultaneous detection of multiple respiratory pathogens, but limited data is available on the comparative diagnostic performance of different testing systems. In this multicenter prospective study, we aimed to compare the QIAstat-Dx Respiratory Panel 2.0 (QIAstat-Dx-RP2.0) with the widely used BioFire-RP2.1, using 269 respiratory clinical specimens. Concordant test results were obtained in 232 (86.3%) cases. Discordant test results included 33 BioFire-RP2.1(+)/QIAstat-Dx-RP2.0(-) and 4 BioFire-RP2.1(-)/QIAstat-Dx-RP2.0(+) results. Discordant samples showed significantly lower pathogen loads than concordant ones (p < 0.01). Overall, the QIAstat-Dx-RP2.0 showed an analytical sensitivity of 50%-100% depending on the respiratory target, with an analyt
SUBMITTER: Gosert R
PROVIDER: S-EPMC11664495 | biostudies-literature | 2024 Dec
REPOSITORIES: biostudies-literature