ABSTRACT: Background
Inactivation of infectious liquid waste can be performed by different means, including autoclaving or chemical inactivation. Autoclaving is most widely used, but cannot always be implemented, so that chemical inactivation is a possible alternative. However, its efficacy has to be proven by in-house validation. Here, we provide an easy-to-implement validation process adapted from existing standards.Methods
The bactericidal activity of two commercially available disinfectants, containing glucoprotamine or peracetic acid, was tested on two gram-positive bacteria, methicillin-resistant Staphylococcus aureus (MRSA) and Streptococcus pneumoniae (SP). Quantitative suspension tests were adapted from the European standard EN 13727. Membrane filtration was us