To improve liver proton density fat fraction (PDFF) and R 2 * $$ {R}_2^{\ast } $$ quantification at 0.55 T by systematically validating the acquisition parameter choices and investigating the performance of locally low-rank denoising methods.Methods
A Monte Carlo simulation was conducted to design a protocol for PDFF and R 2 * $$ {R}_2^{\ast } $$ mapping at 0.55 T. Using this proposed protocol, we investigated the performance of robust locally low-rank (RLLR) and random matrix theory (RMT) denoising. In a reference phantom, we assessed quantification accuracy (concordance correlation coefficient [ ρ c $$ {\rho}_c $$ ] vs. reference values) and precision (using SD) across scan repetitions. We performed in vivo liver scans (11 subjects) and used regions of