Large improvement in RF magnetic fields and imaging SNR with whole-head high-permittivity slurry helmet for human-brain MRI applications at 7 T.


ABSTRACT:

Purpose

To optimize the design and demonstrate the integration of a helmet-shaped container filled with a high-permittivity material (HPM) slurry with RF head coil arrays to improve RF coil sensitivity and SNR for human-brain proton MRI.

Methods

RF reception magnetic fields ( B 1 - $$ {\mathrm{B}}_1^{-} $$ ) of a 32-channel receive-only coil array with various geometries and permittivity values of HPM slurry helmet are calculated with electromagnetic simulation at 7 T. A 16-channel transmit-only coil array, a 32-channel receive-only coil array, and a 2-piece HPM slurry helmet were constructed and assembled. RF transmission magnetic field ( B 1 + $$ {\mathrm{B}}_1^{+} $$ ), B 1 - $$ {\mathrm{B}}_1^{-} $$ , and MRI SNR maps from the entire human brain were measured

SUBMITTER: Soon SH 

PROVIDER: S-EPMC11680740 | biostudies-literature | 2025 Mar