Impact of device variability and protocol differences on kidney function during normothermic machine perfusion: A comparative study using porcine and human kidneys.


ABSTRACT:

Introduction

A growing interest in renal normothermic machine perfusion (NMP) has resulted in more clinically available perfusion devices. While all perfusion systems have the same aim, there are significant differences in their circuits, pumps, sensors, and software. Therefore, our objective was to assess the impact of different perfusion protocols and devices on kidney function and perfusion parameters during NMP.

Methods

Porcine kidneys were subjected to 30 min of warm ischemia, 24 h of static cold storage, and subsequently perfused for 6 h using (1) the Kidney Assist (KA) machine with a pressure of 75 mm Hg, (2) the KA device incorporating several adjustments and a pressure of 85 mm Hg (modified KA), or (3) the Perlife (PL) perfusion device (n = 4). Consecutively, discar