ABSTRACT: Background
Photon-counting detector (PCD) technology has the potential to reduce noise in computed tomography (CT). This study aimed to carry out a voxelwise noise characterization for a clinical PCD-CT scanner with a model-based iterative reconstruction algorithm (QIR).Methods
Forty repeated axial acquisitions (tube voltage 120 kV, tube load 200 mAs, slice thickness 0.4 mm) of a homogeneous water phantom and CTP404 module (Catphan-504) were performed. Water phantom acquisitions were also performed on a conventional energy-integrating detector (EID) scanner with a sinogram/image-based iterative reconstruction algorithm, using similar acquisition/reconstruction parameters. For smooth/sharp kernels, filtered back projection (FBP)- and iterative-reconstructed images were obtai