Unknown

Dataset Information

Mitral valve repair with the semi-rigid Memo 4D annuloplasty ring: early clinical and echocardiographic outcomes from the MANTRA study.


ABSTRACT:

Objectives

Memo 4D is a semi-rigid ring with an exclusive saddle shape and progressive increased anteroposterior diameter. This preliminary analysis reports 30-day clinical and haemodynamic outcomes of the MANTRA Memo 4D sub-study.

Methods

MANTRA is an 'umbrella' prospective, multicentre, worldwide post-market study to collect real-life safety and performance data on the Corcym devices. Clinical and echocardiographic outcomes were gathered preoperatively, at discharge and each follow-up. KCCQ-12 questionnaires were collected preoperatively and at 30 days. Echocardiographic studies were performed per a predefined protocol and assessed by an independent core laboratory.

Results

In total, 166 patients (52, 31.3% female, mean age 60.7 ± 11.4 years) underwent mitral valve

SUBMITTER: Dzemali O 

PROVIDER: S-EPMC11700519 | biostudies-literature | 2024 Dec

REPOSITORIES: biostudies-literature