Quantitative fluorescent nanoparticle tracking analysis and nano-flow cytometry enable advanced characterization of single extracellular vesicles.
Ontology highlight
ABSTRACT: Current state-of-the-art tools for analysing extracellular vesicles (EVs) offer either highly sensitive but unidimensional bulk measurements of EV components, or high-resolution multiparametric single-particle analyses which lack standardization and appropriate reference materials. This limits the accuracy of the assessment of marker abundance and overall marker distribution amongst individual EVs, and finally, the understanding of true EV heterogeneity. In this study, we aimed to define the standardized operating procedures and reference material for fluorescent characterization of EVs with two commonly used EV analytical platforms-nanoparticle tracking analysis (NTA) and nano-flow cytometry (nFCM). We achieved quantitative fluorescence analyses on ZetaView NTA and NanoAnalyzer nFCM instr
SUBMITTER: Mladenovic D
PROVIDER: S-EPMC11707551 | biostudies-literature | 2025 Jan
REPOSITORIES: biostudies-literature
ACCESS DATA