Lucerne milestone approach for benchmarking and education: Towards ultra-low dose endovascular aortic repair.


ABSTRACT:

Objective

The aim of this single-center case series is to demonstrate that an ultra-low dose (ULD) can be routinely achieved in the hybrid operating room in standard endovascular aortic repair (EVAR) for infrarenal abdominal aortic aneurysm by adjusting the manufacturer's predefined imaging parameters, hardware configurations and user protocols (including benchmarking).

Methods

The hybrid operating room manufacturer predefined EVAR software setup of the dose exposure control software (OPTIQ, Siemens Healthineers, Forchheim, Germany) at our university medical center was screened for possible improvements regarding radiation dose application. Tests on a water-equivalent as well as polymethyl methacrylate phantom model to assess the impact of technical settings were performed,