Bone imaging modality precision and agreement between DXA, pQCT, and HR-pQCT.


ABSTRACT: Quantifying precision error for DXA, peripheral QCT (pQCT), and HR-pQCT is crucial for monitoring longitudinal changes in body composition and musculoskeletal outcomes. Agreement and associations between bone variables assessed using pQCT and second-generation HR-pQCT are unclear. This study aimed to determine the precision of, and agreement and associations between, bone variables assessed via DXA, pQCT, and second-generation HR-pQCT. Thirty older adults (mean age: 64.2 ± 8.0 yr; women: 67%) were recruited. DXA scans were performed at the total hip, lumbar spine, and whole body. Distal (4%) and proximal (30%/33%/66%) skeletal sites at the radius and tibia were scanned with pQCT and/or HR-pQCT. Root-mean-squared coefficients of variation (%CVRMS) were calculated to define

SUBMITTER: Mesinovic J 

PROVIDER: S-EPMC11752644 | biostudies-literature | 2025 Feb