ABSTRACT: Purpose
To evaluate CT Lucia 602 intraocular lenses implanted in cases lacking capsular support and explanted because of decentration/dislocation with subsequent pathological analysis. The main objectives were to (1) document the status of haptics and (2) verify the presence of the rotisserie effect.Setting
John A. Moran Eye Center, University of Utah, Salt Lake City, Utah.Design
Case series with clinicopathological correlation.Methods
13 explanted CT Lucia 602 lenses were included: 12 implanted by the Yamane technique and 1 that was iris-fixated. 4 lenses were explanted between October 2020 and February 2022. 9 lenses, explanted between November 2022 and February 2024, also exhibited severe optic tilt, leading to explantation. A chart review was performed t