Multi-center benchmarking of cervical spinal cord RF coils for 7 T MRI: A traveling spines study.


ABSTRACT:

Purpose

The depth within the body, small diameter, long length, and varying tissue surrounding the spinal cord impose specific considerations when designing radiofrequency coils. The optimal coil configuration for 7 T cervical spinal cord MRI is unknown and, currently, there are very few coil options. The purpose of this work was (1) to establish a quality control protocol for evaluating 7 T cervical spinal cord coils and (2) to use that protocol to evaluate the performance of 4 different coil designs.

Methods

Three healthy volunteers and a custom anthropomorphic phantom (the traveling spines cohort) were scanned at seven 7 T imaging centers using a common protocol and each center's specific cervical spinal cord coil. Four different coil designs were tested (two in-house, on

SUBMITTER: Alonso-Ortiz E 

PROVIDER: S-EPMC11761089 | biostudies-literature | 2025 Jan

REPOSITORIES: biostudies-literature