Reusable Biosensor for Easy RNA Detection from Unfiltered Saliva.


ABSTRACT: Biosensors are transforming point-of-care diagnostics by simplifying the detection process and enabling rapid, accurate testing. This study introduces a novel, reusable biosensor designed for direct viral RNA detection from unfiltered saliva, targeting SARS-CoV-2. Unlike conventional methods requiring filtration, our biosensor leverages a unique electrode design that prevents interference from saliva debris, allowing precise measurements. The biosensor is based on electrochemical principles, employing oligonucleotide probes immobilized on a hydrophobic-coated electrode, which prevents air bubbles and salt crystal formation. During validation, the biosensor demonstrated a sensitivity and specificity of 100%, accurately identifying SARS-CoV-2 in saliva samples without false positives or nega

SUBMITTER: Wityk P 

PROVIDER: S-EPMC11769206 | biostudies-literature | 2025 Jan