ABSTRACT: Significant developments have been made in the field of wearable healthcare by utilizing soft materials for the construction of electronic sensors. However, the lack of adaptability to complex topologies, such as ear canal, results in inadequate sensing performance. Here, we report an in-ear physiological sensor with mechanical adaptability, which softens upon contact with the ear canal's skin, thus reducing the sensor-skin mechanical mismatch and interface impedance. An efficient strategy of mechanical adjustment and switching is exploited to increase the softness of the device, leading to a significant decrease in Young's modulus from 30.5 MPa of thermoplastic polyurethane (TPU) to 0.86 MPa of TPU/Ecoflex foam (TEF).The mechanical adaptability at body temperature endows the in-ear device