ABSTRACT: Purpose
Serum caffeine concentration is an indicator of caffeine intoxication; however, it is difficult to measure it in most emergency departments. We developed a simple estimation method using a point-of-care test kit for urinary caffeine.Methods
Caffeine-spiked human serum (100, 50, 25, and 10 µg/mL) was diluted 10-, 20-, 50-, and 100-fold with phosphate-buffered saline and applied to the kit. After 5 min incubation, the kit was scanned by a flatbed scanner and the membrane image was processed with ImageJ.Results
When the 20-fold diluted serum was applied, serum samples with initial caffeine concentration ≤ 25 and ≥ 50 µg/mL were caffeine-negative and -positive, respectively. When the 100-fold diluted serum was applied, none of the caffeine-spiked serum samples g