Ontology highlight
ABSTRACT: Purpose
To determine the feasibility of simultaneous multi-slice (SMS) real-time MRI (RT-MRI) at 0.55T for the evaluation of cardiac function.Methods
Cardiac CINE MRI is routinely used to evaluate left-ventricular (LV) function. The standard is sequential multi-slice balanced SSFP (bSSFP) over a stack of short-axis slices using electrocardiogram (ECG) gating and breath-holds. SMS has been used in CINE imaging to reduce the number of breath-holds by a factor of 2-4 at 1.5T, 3T, and recently at 0.55T. This work aims to determine if SMS is similarly effective in the RT-MRI evaluation of cardiac function. We used an SMS bSSFP pulse sequence with golden-angle spirals at 0.55T with an SMS factor of three. We cover the LV with three acquisitions for SMS, and nine for single-band (
SUBMITTER: Yagiz E
PROVIDER: S-EPMC11782716 | biostudies-literature | 2025 Apr
REPOSITORIES: biostudies-literature