Radiofrequency-transparent local B<sub>0</sub> shimming coils using float traps.


ABSTRACT:

Purpose

Static field (B0) inhomogeneities present a major challenge in high-field MRI. Multicoil shimming using independent, local, direct-current (DC) shim coils has emerged as a powerful and flexible technique to address this issue. However, many-turn DC coils can lead to significant mutual coupling with radiofrequency (RF) coils, causing transmit field (B1 +) distortions and signal-to-noise ratio degradation.

Methods

We introduce an innovative RF-transparent DC coil that performs B0 shimming while minimizing RF performance impact. The design incorporates float traps to maintain high RF impedance, allowing flexible placement relative to the RF coil without compromising signal-to-noise ratio or affecting B1 +