ABSTRACT: Background
Metasurface coils (MCs) are a promising magnetic resonance imaging (MRI) technology. Aiming to evaluate the image quality of MCs for knee and elbow imaging, we compared signal-to-noise ratio (SNRs) obtained in standard clinical setups.Methods
Knee and elbow MRI routine sequences were applied at 1.5 T, implementing four coil scenarios: (1) 15-channel transmit/receive knee coil; (2) four-channel multipurpose coil (flex coil); (3) MC + spine coil; and (4) MC + multipurpose coil. Three regions of interest (ROIs) at different anatomical depths were compared.Results
Seven participants (aged 28 ± 2 years; 6 males) were enrolled. In elbow MRI, the MC + spine coil demonstrated the highest SNR across all ROIs (superficial-anterior: +114%, p = 0.008; middle: +147%,