ABSTRACT: Background
Intracardiac echocardiography (ICE) is an alternative to transesophageal echocardiography to guide left atrial appendage occlusion (LAAO). However, 2-dimensional ICE has important limitations that hinder its adoption in routine practice.Objectives
This prospective multicenter study (NCT04569734) investigated the feasibility and safety of ICE-guided LAAO with Watchman FLX using a novel 3-dimensional ICE catheter (VeriSight Pro, Philips).Methods
A total of 100 patients undergoing LAAO with the Watchman FLX device were recruited. All cases were performed under moderate sedation. A simplified 2 orthogonal view ICE imaging protocol was adopted to assess PASS (position, anchoring, sizing, and sealing) criteria. The feasibility endpoint was successful implantation of the FLX device and adequate seal (peri-device leak <5 mm) at 45 days. The safety endpoint was the incidence of major complications at 7 and 45 days. Physician experience and imaging quality scores were collected (1 = lowest, 5 = highest).Results
The mean age was 77.0 ± 8.1 years, and 39 (39%) were women. CHA2DS2-VASc score was 4.8 ± 1.5 and HAS-BLED score was 3.2 ± 1.1. Total procedural time was 54 ± 25 minutes. Mean contrast volume was 41 ± 46 ml, with no contrast used in 16 patients. The feasibility endpoint was achieved in 95 patients (95%); 4 patients were not implanted due to unsuitable anatomy, and 1 patient had a leak >5 mm at 45 days. There was 1 periprocedural stroke and 2 gastrointestinal bleeds requiring transfusion at 7 days with no other major complications. Physicians' rating of overall ICE imaging quality was 4.6 ± 0.6.Conclusions
The novel 3-dimensional ICE VeriSight Pro Probe can safely and effectively guide LAAO with Watchman FLX using a simplified imaging protocol.
SUBMITTER: Sularz A
PROVIDER: S-EPMC11782859 | biostudies-literature | 2025 Feb
REPOSITORIES: biostudies-literature

JACC. Advances 20250113 2
<h4>Background</h4>Intracardiac echocardiography (ICE) is an alternative to transesophageal echocardiography to guide left atrial appendage occlusion (LAAO). However, 2-dimensional ICE has important limitations that hinder its adoption in routine practice.<h4>Objectives</h4>This prospective multicenter study (NCT04569734) investigated the feasibility and safety of ICE-guided LAAO with Watchman FLX using a novel 3-dimensional ICE catheter (VeriSight Pro, Philips).<h4>Methods</h4>A total of 100 pa ...[more]