Comparative evaluation of sensititre YeastOne and VITEK2 antifungal susceptibility tests with CLSI broth microdilution method of clinical <i>Cryptococcus</i> isolates in Taiwan.


ABSTRACT: Commercial antifungal susceptibility tests were available for clinical yeast isolates. However, the updated Sensititre YeastOne (SYO) version YO10C excluded Cryptococcus species for susceptibility testing. Uncorrelation of antifungal susceptibility patterns by SYO and therapeutic outcomes had been recently reported. We compared the performance of current commercial susceptibility tests with the standard CLSI broth microdilution (BMD) method for clinical Cryptococcus isolates. Forty-seven clinical Cryptococcus isolates were included from 1 January 2012 to 30 June 2023, among which 44 isolates were Cryptococcus neoformans while 3 were Cryptococcus gattii. The performance of SYO version YO10C and VITEK2 YS09 was compared with the CLSI BMD method and correlat

SUBMITTER: Weng T-P 

PROVIDER: S-EPMC11792495 | biostudies-literature | 2025 Feb