ABSTRACT: Background
The performance of continuous glucose monitoring (CGM) systems is difficult to compare due to different study designs and a lack of head-to-head studies. This study evaluated the performance of FreeStyle Libre 3 (FL3), Dexcom G7 (DG7), and Medtronic Simplera (MSP) against different comparator methods and during clinically relevant glycemic scenarios.Method
Twenty-four adult participants with type 1 diabetes mellitus wore one sensor of each CGM system in parallel for up to 15 days. Sensors of DG7 and MSP were exchanged on days 5 and 8, respectively. Three 7-hour sessions with 15-minute comparator blood glucose-level measurements using YSI 2300 (YSI, venous), Cobas Integra (INT, venous), and Contour Next (CNX, capillary) were conducted on days 2, 5, and 15. Simulta