Ontology highlight
ABSTRACT: Background
Implantable cardioverter defibrillators (ICD) are battery-operated devices used to manage irregular heart rhythms and deliver therapeutic shocks to the heart. This updated systematic review and meta-analysis compares the efficacy and extraction-related outcomes of single-coil versus dual-coil ICDs in view of conflicting data.Methods
Several databases, including PubMed, Cochrane Library, and Google Scholar, were comprehensively explored dating from inception to April 1, 2024. Data were compared using odds ratio (OR), hazard ratio (HR), and mean differences (MD). A value of p < 0.05 indicated statistical significance.Results
Ultimately, 28 studies were included in this quantitative synthesis. Defibrillation threshold (DFT) indicated statistical superiority
SUBMITTER: Tahir MF
PROVIDER: S-EPMC11799043 | biostudies-literature | 2025 Feb
REPOSITORIES: biostudies-literature