Unknown

Dataset Information

A dosimetric comparison of non-coplanar volumetric modulated arc therapy and non-coplanar fixed field intensity modulated radiation therapy in hippocampus-avoidance whole-brain radiation therapy with a simultaneous integrated boost for brain metastases.


ABSTRACT:

Objective

The aim of this study was to investigate the dosimetric differences between non-coplanar volumetric modulated arc therapy (VMAT) and non-coplanar fixed-field intensity-modulated radiotherapy (IMRT) in hippocampus-avoidance whole-brain radiation therapy with a simultaneous integrated boost (HA-WBRT+SIB) for brain metastases using the Monaco treatment planning system (TPS).

Method

A total of 22 patients with brain metastases were retrospectively enrolled. Two radiotherapy treatment plans were designed for each patient: non-coplanar VMAT and non-coplanar fixed field IMRT. The dose distribution of targets and organs at risk (OAR), the number of monitor units (MUs), and pre-treatment plan verification were compared between the two plans while meeting the prescribed dose

SUBMITTER: Zeng H 

PROVIDER: S-EPMC11799273 | biostudies-literature | 2024

REPOSITORIES: biostudies-literature