In-clinic vitreous biopsy peel pack technique.


ABSTRACT:

Background

Vitreous biopsy is a common technique used to guide management of acute endophthalmitis and help differentiate between infectious and inflammatory conditions. Currently, in-clinic vitreous biopsy is performed with a 25-gauge needle, without the ability to cut vitreous, potentially leading to reduced diagnostic yield. Recent work demonstrated the ability to perform vitreous biopsy with an off-the-shelf vitreous cutter. However, this was limited by complexity of assembly. Here, a technique using a single peel pack vitrectomy cutter is demonstrated for in-clinic vitreous biopsy.

Methods

A 25-gauge vitreous cutter is opened from a peel pack. The drive line is identified, cut to length, and attached to a 10 mL syringe. A 1 mL syringe is attached to the aspiration line.