Utilizing fundus images captured by two ultra-wide field imaging systems to measure diagnostic indicators and assess the grade of diabetic retinopathy.


ABSTRACT:

Background

This study compared the effectiveness of different ultrawide field fundus imaging systems (Clarus™ 500 and Optos®) in diagnosing diabetic retinopathy (DR).

Methods

This was a prospective, multicentre study. Retinal photographs were captured at four eye centres utilizing both the Clarus™ 500 and Optos® imaging systems. The image quality and the effective retinal area were compared. Then the consistency in diagnosing the severity of DR by two distinct imaging systems was compared according to three separate grading criteria (the International Council of Ophthalmology, the Early Treatment Diabetic Retinopathy Study [ETDRS] grading diagnostic criteria, and the Chinese Clinical Diagnosis and Treatment Guidelines for Diabetic Retinopathy [2014]) we