ABSTRACT: Purpose
To evaluate the effects of axial length and slot-to-slot distance of double-slot microwave antenna (DSMA) with frequency of 433 MHz on the size and shape of ablation zones created under different input microwave powers.Materials and methods
The design of double slot microwave antennas (DSMAs) with axial lengths (70 mm, 30 mm) and slot-to-slot distance (49 mm, 10 mm) were optimized by numerical simulation and ex vivo liver experiments. Finite-element method simulations and forty ablations of swine liver were employed to obtain the temperature distributions within liver tissue using DSMAs at the 433 MHz operating frequency in a range of heating powers (20, 30, 40 and 50W) for 600 s. The dependence of the effectiveness of MWA on the axial length and slot-to-slot distan