A Comparative Analysis of In Vivo-Generated and Artificial CoCrMo Wear Particles Created by High-Energy Ball Milling and the Buchhorn Method.
Ontology highlight
ABSTRACT: This study compares CoCrMo particles generated by a high-energy ball milling method with those produced by the established Buchhorn method and with in vivo-generated wear particles from periprosthetic tissue. CoCrMo particles were produced utilization both methods. The particle size distribution was analyzed using laser diffraction, and the morphology was examined through scanning electron microscopy (SEM). Energy-dispersive X-ray spectroscopy (EDS) provided a qualitative analysis of the material composition. The high-energy ball milling method yielded CoCrMo particles with a D50 of 7.92 µm, a significantly smaller value than the D50 of 88.30 µm observed for Buchhorn particles. The SEM analysis demonstrated that the particles produced by the ball mill have a roundish, red blood cell-like a