ABSTRACT: Objectives
To determine the appropriate intraosseous (IO) needle insertion site, optimal depth and success using a drill-assisted device (DAD) versus a manually inserted needle (MIN).Methods
Computed tomography scans of neonatal cadavers were analyzed. Success was based on tibial needle tip placement within the marrow cavity and contrast media distribution.Results
Nineteen cadavers (38 tibiae) were included. The overall success rate was comparable between DAD and MIN needles, but reduced in very-low birthweight (VLBW) infants. The insertion site was consistent across birth weight groups. Contrast leakage occurred overall in 15.8% and 41.7% in VLBW infants and was insignificantly greater in DAD versus MIN needles. Minimum and maximum puncture depth was adjusted for h