ABSTRACT: Background
Malposition of the acetabular component during total hip arthroplasty (THA) is a leading cause of complications and need for revision. Robotic-assisted THA purports to improve accuracy of component positioning with many reports demonstrating over 92% of components within 10° of inclination and anteversion compared to intraoperative system output. This study aimed to evaluate the intraoperative accuracy output of acetabular cup position values using a handheld miniaturized portable navigation system (Naviswiss, AG) compared to postoperative computed tomography (CT) scans.Methods
A total of 108 direct anterior approach THA surgeries using the intraoperative navigation device were performed over a 6-month period. Intraoperative device output for measured acetabular