Challenging the skin pigmentation bias in tissue oximetry via time-domain near-infrared spectroscopy.
Ontology highlight
ABSTRACT: Recently, skin pigmentation has been shown to affect the performance of pulse oximeters and other light-based techniques like photo-acoustic imaging, tissue oximetry, and continuous wave near-infrared spectroscopy. Evaluating the robustness to changes in skin pigmentation is therefore essential for the proper use of optical technologies in the clinical scenario. We conducted systematic time-domain near-infrared spectroscopy measurements on calibrated tissue phantoms and in vivo on volunteers during static and dynamic (i.e., arterial occlusion) measurements. To simulate varying melanosome volume fractions in the skin, we inserted, between the target sample and the measurement probe, thin tissue phantoms made of silicone and nigrosine (skin phantoms). Additionally, we conducted an extensive
SUBMITTER: Lacerenza M
PROVIDER: S-EPMC11828448 | biostudies-literature | 2025 Feb
REPOSITORIES: biostudies-literature
ACCESS DATA