Intraoperative radiotherapy IORT applicators for treatment of small skin lesions a phantom and planning study.


ABSTRACT: Radiation therapy is actively utilized for superficial lesions. External beam radiotherapy for cutaneous lesions utilizes electrons with low transmittance. Conventional electron beam therapy uses Cerrobend blocks for field shaping; however, the Intraoperative Radiotherapy (IORT) applicator offers superior dosimetric characteristics. The dosimetric parameters were measured using 4 and 6 MeV electron beams delivered by a Trilogy linear accelerator, and percent depth dose and lateral dose profiles were compared under the presence of the IORT applicator and Cerrobend block. The dose calculations under various IORT applicator conditions and planning studies were performed using Monte Carlo simulation. Treatment plans for three sites were evaluated in terms of coverage of the planning target vol

SUBMITTER: Lee US 

PROVIDER: S-EPMC11829022 | biostudies-literature | 2025 Feb