Unknown

Dataset Information

Innovative balloon-inflatable venous cannula for enhanced cardiopulmonary bypass in minimally invasive cardiac surgery.


ABSTRACT:

Objectives

Minimally invasive cardiac surgery (MICS) struggles with effective caval isolation and cannulation for cardiopulmonary bypass (CPB). We aimed to develop a novel MICS venous cannula, eliminating external manipulations. The goal of this study is to thoroughly evaluate both the safety and the efficacy of a newly developed venous cannula.

Methods

The study presents the Aulus venous cannula for MICS, designed with internal balloons to block caval blood flow. Preclinical bench tests with a cardiac biosimulator and large animal studies per ISO10993-2016, evaluated performance and safety.

Results

The heparin-coated Aulus cannula had a post-sterilization comparable density of ∼0.200 μg/cm2. In ex vivo tests, using porcine heart models, the cannula enabled full caval

SUBMITTER: Sazzad F 

PROVIDER: S-EPMC11829808 | biostudies-literature | 2025 Feb

REPOSITORIES: biostudies-literature