ABSTRACT: Objectives
Percutaneous balloon mitral valvuloplasty (PBMV) for rheumatic mitral stenosis (MS) using the classical Inoue technique can be technically challenging, especially in difficult anatomies. This paper describes a novel wire-based technique of PBMV that involves advancement of the Inoue balloon over a preshaped stiff 0.035-inch wire.Key steps
These include targeted transseptal puncture, advancement of a steerable sheath to the left atrium, crossing the diseased mitral valve, exchanging for a 0.035-inch preshaped wire in the left ventricle, slenderization of the appropriately sized Inoue balloon on the wire, advancement and positioning of the balloon across the mitral valve, and balloon inflation.Potential pitfalls
This technique does use additional equipment,
SUBMITTER: Gupta T
PROVIDER: S-EPMC11830263 | biostudies-literature | 2025 Feb