Ontology highlight
ABSTRACT: Background
The Rapid-O2 oxygen insufflation device® (Rapid-O2) was designed primarily for rescue oxygenation in cannot intubate, cannot oxygenate (CICO) events; thus, hypercapnia is inevitable. Rapid-O2 was modified to enhance ventilation using the Venturi effect during expiration.Methods
To determine the most effective combination of inner catheters (20 gauge [G], 18 G, 16 G, 14 G, and 2-mm inner diameter [ID] transtracheal catheter [TTC]) and insufflation catheters (16 G, 14 G, and 2-mm ID TTC) for achieving optimum ventilation, insufflating and expiratory flows were measured at an oxygen flow rate of 15 L/min. The insufflating and expiratory pressures were measured at 6-15 L/min. The flows and pressures were measured using a gas flow analyzer. The insufflating and expira
SUBMITTER: Jang J
PROVIDER: S-EPMC11836609 | biostudies-literature | 2025 Feb
REPOSITORIES: biostudies-literature