ABSTRACT: Objective
Phenox flow diverters (p64 and its smaller vessel variant p48) represent an established treatment option for intracranial aneurysms. This study evaluates the safety and efficacy of the new generation of these devices with an additional antithrombotic surface coating (HPC).Methods
Consecutive patients treated between 2020 and 2023 at three institutions were retrospectively reviewed for aneurysm characteristics, procedural details, complications, and angiographic outcomes.Results
Sixty-one patients (mean age 56 years) were treated for 61 aneurysms. The mean aneurysm size was 8.3 ± 4.9 mm, 12 (19.7%) were ruptured, 16 (26.2%) were recurrent after previous treatment, 10 (16.4%) were located in the posterior circulation, and 7 (11.5%) had nonsaccular morphology