ABSTRACT: Objectives
Ventilation of bearded patients using the standard face mask (FM) is often difficult, particularly in field settings and mass casualty events. The current study compares the effectiveness of a novel Bag-Valve-Guedel Adaptor (BVGA) with the standard FM when applied to anesthetized patients by anesthesiology trainees with limited experience.Methods
Male patients scheduled for elective surgery (American Society of Anesthesiology physical score 1-2) were recruited for this prospective, randomized, cross-over trial. Beard length was categorized as <0.5 cm (none/stubble), 0.5 to 1 cm, 1 to 5 cm, or ≥5 cm. Anesthetized patients were ventilated by anesthesiology trainees using the BVGA and an FM. The main outcome included end-tidal CO2, expiratory tidal volume