Unknown

Dataset Information

Measurement accuracy and cutoffs for predicting primary aldosteronism diagnosis using Lumipulse® for renin and aldosterone measurements.


ABSTRACT:

Objectives

Primary aldosteronism (PA) diagnosis is a multistep process that begins with screening based on the plasma aldosterone-to-renin ratio (ARR). As cut-off values depend on the methods used, the aim of this study was 1) to validate the performances of new aldosterone and renin chemiluminescent immuno-assays using the Lumipulse® analyzer (Fujirebio®) and then 2) to define the criteria to distinguish PA patients from those with essential hypertension (EH).

Methods

This monocentric retrospective study included 297 patients admitted to Georges Pompidou European Hospital, France, between January 2021 and December 2022 for the assessment of hypertension etiology. We evaluated analytical performances and compared Lumipulse® results for plasma renin (PRC) and plasma aldostero

SUBMITTER: Laney M 

PROVIDER: S-EPMC11856259 | biostudies-literature | 2025

REPOSITORIES: biostudies-literature