Unknown

Dataset Information

Development of a Tool for Verifying Leakage Detection in Microfluidic Systems.


ABSTRACT: While submissions of microfluidic-based medical devices to the Food and Drug Administration (FDA) have increased in recent years, leakage remains a common but difficult failure mode to detect in microfluidic systems. Here, we have developed a sensitive tool to measure and verify leakages ranging from 0.1% to 10% in leakage detection systems, which can then be used to detect leak in microfluidic devices. Our methodology includes an analytical model that applies hydrodynamic resistance using different fluid-contacting elements (e.g., tubing, junctions, and connectors) to tune the leakage rate based on the application-specific acceptance criteria. We then used three polymer-based microfluidic systems to target leakage rates of approximately 0.1, 1.0, and 10%. The experimental uncertainties in

SUBMITTER: Bozorgnezhad A 

PROVIDER: S-EPMC11857335 | biostudies-literature | 2025 Jan

REPOSITORIES: biostudies-literature