A Systematic Study of Liquid Chromatography in Search of the Best Separation of Cannabinoids for Potency Testing of Hemp-Based Products.
Ontology highlight
ABSTRACT: A study was conducted to search for the best separation of eighteen cannabinoids, the maximum number of cannabinoids that have been quantified so far, for potency testing of hemp-based products using liquid chromatography diode array detector (LC-DAD). The investigation utilized four column types, all sharing the same dimension (150 mm × 2.1 mm) and core-shell particle size (2.7 µm), but different stationary phases: dimethyl-octadecyl (Poroshell 120 EC-C18), diisobutyl-octadecyl (Raptor ARC-18), reverse phase (RP)-carbamate (Cortecs Shield RP-18), and RP-amide (Ascentis Express RP-Amide). The resolution of adjacent cannabinoids was kept close to 1.5 or higher, while the separation time was kept as short as possible. The fastest separation was achieved within 15.0 min using two sequential R
SUBMITTER: Owolabi A
PROVIDER: S-EPMC11858520 | biostudies-literature | 2025 Feb
REPOSITORIES: biostudies-literature