Analysis of the optical performance of intraocular lenses using profilometric measurements.


ABSTRACT:

Purpose

The aim of this study was to develop a methodology, based on profilometer measurements to assess the optical behaviour of Intraocular Lenses (IOls). The "Modulation Transfer Function through-object" (MTF through-object) based on vergence object displacement was calculated for different pupil sizes and pseudophakic eyes. Tilt and decentration were also analysed in a realistic cornea eye model.

Methods

For comparison between the different IOLs, an optical quality criterion based on a minimum value the MTF through-object and the recognition of simulated vision optotypes was introduced. Five IOLs were used in this study: Tecnis Eyhance, Mini Well, Tecnis Symfony, Tecnis Synergy and RayOne EMV.

Results