Utility of under-sampled scans with iterative reconstruction and high-frequency preserving transform for high spatial resolution magnetic resonance cholangiopancreatography.


ABSTRACT:

Purpose

Under-sampled scans with iterative reconstruction and high-frequency preserving transform (Us-IRHF) can increase the acquisition speed without degrading the image quality by recovering image information from under-sampled data. We investigate the clinical applicability of high spatial resolution magnetic resonance cholangiopancreatography (MRCP) images without extending the scanning time using Us-IRHF.

Methods

A slit phantom was scanned with conventional- (without Us-IRHF), Us-IR- (without HF), and Us-IRHF scanning. The matrix size was 320 × 320 for Us-IR- and Us-IRHF- and 288 × 208 for conventional scanning. Modulation transfer function (MTF) focused on the 1.0 lp/cm gauge for each scanning was calculated. For clinical study we acquired respiratory-triggered 3D MRCP