Analysis of urinary organic acids (UOAs) by gas chromatography mass-spectrometry (GC-MS) is widely used in metabolomic studies. It is a complex test with many limitations and pitfalls yet there is limited evidence in the literature to support best practice. This study investigated the impact of drying down time and temperature on the recovery of 16 key analytes from solvent extracts.Methods
Pooled urine specimens were enriched with organic acids. Urine aliquots (n = 3) were acidified and extracted into diethyl ether and ethyl acetate. Extracts were dried under nitrogen at ambient temperature (25 °C); 40 °C; 60 °C then left for 0; +5; +15 min. Dried extracts were derivatised with N,O,-bis-(trimethylsilyl)trifluoroacetamide prior to analysis by GC-MS. Urine specime