ABSTRACT: Objective
Magnetic resonance (MR) relaxometry is an absolute and reproducible quantitative method, compared with signal intensity for the evaluation of liver biliary function. This is obtainable by the T1 reduction rate (T1RR), as it carries a smaller systematic error than the pre/post contrast agent T1 measurement. We aimed to develop and test an MR T1 relaxometry tool tailored for the evaluation of liver T1RR after gadolinium ethoxybenzyl-diethylenetriaminepentaacetic acid administration on 1.5T MR.Methods
In vitro/vivo (liver) T1RR values with two 3D FLASH variable-flip-angle sequences were calculated by a MATLAB algorithm. In vitro measurements were done by 2 physicists, in consensus. The prospective in vivo study was approved by the local ethical committee and performe