Wearable ECG patch monitoring for 72 h is comparable to conventional Holter monitoring for 24 h to detect cardiogenic vertigo.
Ontology highlight
ABSTRACT: We sought to compare the diagnostic efficacy of a 72-hour wearable electrocardiography (ECG) patch monitoring with a conventional 24-hour Holter monitoring for the detection of cardiogenic vertigo. We conducted a prospective multicenter study that recruited 47 patients suspected of cardiogenic vertigo in four referral-based university hospitals between November 2019 and April 2022. Patients received simultaneous ambulatory ECG recordings using a 72-hour wearable ECG patch monitoring and a conventional 24-hour Holter monitor. The primary outcome was the detection rate of arrhythmia events during the entire wearing period. The secondary outcomes included device preference and factors affecting the detection of cardiogenic vertigo. During the 72-hour monitoring period, there was no significan
SUBMITTER: Kim HA
PROVIDER: S-EPMC11883029 | biostudies-literature | 2025 Mar
REPOSITORIES: biostudies-literature