Clinical validation of a novel hand dexterity measurement device.


ABSTRACT: The lack of sensitive objective outcome measures for hand dexterity is a barrier for clinical assessment of neurological conditions and has negatively affected clinical trials. Here, we clinically validate a new method for measuring hand dexterity, a novel hand worn sensor that digitises the Finger Tapping Test. The device was assessed in a cohort of 180 healthy controls and 51 people with Amyotrophic Lateral Sclerosis (ALS) and compared against rating scales and traditional measures (Nine Hole Peg test and grip dynamometry). 14 features were extracted from the device and using a logistic regression algorithm, a 0-100 dexterity performance score was generated for each participant, which accounted for age/sex differences. The device returned objective ratings of a participant's hand dexteri

SUBMITTER: Hayden CD 

PROVIDER: S-EPMC11893126 | biostudies-literature | 2025 Mar

REPOSITORIES: biostudies-literature